Wuxi Tired Instrument Technology Co. Ltd
PRODUCTS
HOT PRODUCT


SHG Double Tray Infrared Dryer
- Commodity name: SHG Double Tray Infrared Dryer
- Commodity ID: b04-002
Description
I. General
This dryer is used for rapid drying when determining the water content of sand samples (original second, type second and core sand). It utilizes infrared bulb radiation heat energy to make the water in the sand samples evaporate rapidly, and then weighs them with balance to calculate the water content of the sand samples. Its structure mainly consists of shell, infrared bulb, timer, switch, indicator and other parts.
II. Technical Parameters
1. Infrared bulb ..............................................250W/220V
2. surface temperature of sand bearing disk ........................................110℃~170℃
3. Size of sanding disk (bottom diameter×depth)........................... 110×27mm
4. Drying time .................................................6~10
5. Timer range ..............................................0~15 minutes
6. Overall dimensions .................................................473×223×280mm
7. Weight .......................................................9.6kg
III. Use and adjustment
1. Pour the weighed sand sample 50±0.01g into the sand disk 10 and spread it evenly. 2.
2. Place the sand tray with the sand sample inside the instrument. 3.
3. Turn on the power switch 4, the indicator light 5 on. 4.
4. According to the time required for drying, respectively adjust the timer screw, infrared light bulb is working. 5.
5. After drying the sand sample (about 6 to 10 minutes), the infrared bulb 6 goes out by itself.
Remove the sand sample and leave it to cool (to prevent moisture absorption), pour it into a glass dish, then weigh it with a balance, accurate to 0.01 grams, and calculate its moisture content.
X = (Q1-Q2 )/Q2 × 100%
X water content of sand samples %%
Q1 Weight of sand sample before drying (g)
Q2 weight of sand sample after drying (g)
6. The dryer at the same time, respectively, drying two sand samples.
IV. Repair and maintenance
1. Pay attention to whether the circuit is insulated.
2. Turn off the power after each day, wipe the dust and sand particles from the instrument and the sand bearing disk. 3.
3. When not in use, cover with plastic film to prevent dust. 4.
4. Do not reverse the timer screw to prevent damage and affect the accuracy of timing.